Z-1994-2021Class II

5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF...

Bard Access Systems, Inc. / initiated 2021-05-19 / Terminated / root cause: manufacturing process

Bard Access Systems Inc. began a recall of 5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm... on . FDA classified it as Class II and gives the reason as "Catheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1994-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
5 Fr IR PowerPICC Convenience Kit Catheter Dual Lumen with 70 cm Guidewire Unit Label: PowerPICC REF CK000340B BARD MYPICC KIT 5F French Size Dual-Lumen 70cm Guidewire
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
Catheter convenience kits contain the incorrect guidewire. Kits were labeled with 70cm guidewires, but contained a 50cm guidewire.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 05/19/2021, the Recalling Firm sent an "URGENT MEDICAL DEVICE CORRECTION" Notification via FedEx 2-day to customers informing them that the Recalling Firm has confirmed through customer complaint that a number of catheter convenience kits labeled as having 70cm guidewires may actually contain 50cm guidewires. The Recalling Firm is providing customers with a quick way of identifying if their kit contains the correct or incorrect size guidewire by determining the number of loops of the wire. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Determine if the product has the correct product based on the number of loops of the guidewire while the product is still in the packaging material. Product may be discarded when clinically necessary and replacement requested. 2. Share this recall notification with all users within or outside their facility network to which they may have further distributed the affected products to ensure these additional users are also aware of this advisory notice. 3. Complete the attached Customer Response Form and return to the Recalling Firm contact noted on the form whether or not they have any of the impacted material so that Recalling Firm may acknowledge receipt of this notification per FDA requirements and subsequently process the product replacement, if necessary. For further assistance or for questions - contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
Quantity in commerce
30 kits
Distribution
U.S. Nationwide distribution in the state of RI.
Product codes and lots
Catalog (REF) Number: CK000340B UDI Code: (01)00801741126697 Lot Number: 20BBK658

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1994-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1994-2021Class IITerminatedVoluntary: Firm initiated