Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled...
Baxter Healthcare Corporation began a recall of Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids on . FDA classified it as Class I and gives the reason as "System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2110-2021
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
- Product code
- FRN Pump, Infusion
- Reason for recall
- System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation. The firm is developing a software patch for all Spectrum IQ infusion pumps.
- Quantity in commerce
- 130135 units
- Distribution
- United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
- Product codes and lots
- All Serial Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2110-2021 | Class I | Terminated | Voluntary: Firm initiated |