Z-2110-2021Class I

Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled...

Baxter Healthcare Corporation / initiated 2021-06-04 / Terminated / root cause: software

Baxter Healthcare Corporation began a recall of Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids on . FDA classified it as Class I and gives the reason as "System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2110-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Product code
FRN Pump, Infusion
Reason for recall
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Root cause tag
software
FDA root cause
Software design
Action
Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation. The firm is developing a software patch for all Spectrum IQ infusion pumps.
Quantity in commerce
130135 units
Distribution
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2110-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2110-2021Class ITerminatedVoluntary: Firm initiated