Z-2051-2021Class II
EASYGRIP FLO-41 US. For delivering hemostatic agents to bleeding sites through a trocar
Baxter Healthcare Corporation / initiated 2021-06-15 / Terminated / root cause: manufacturing process
Baxter Healthcare Corporation began a recall of EASYGRIP FLO-41 US. For delivering hemostatic agents to bleeding sites through a trocar on . FDA classified it as Class II and gives the reason as "There is a discrepancy in the expiry date encoded into the 2D barcode. The correct expiration date is the human-readable date". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2051-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- EASYGRIP FLO-41 US. For delivering hemostatic agents to bleeding sites through a trocar.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- There is a discrepancy in the expiry date encoded into the 2D barcode. The correct expiration date is the human-readable date
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On June 15th 2021, the firm sent a letter to its customers with the following instructions: 1. Healthcare providers may continue to safely use the EASYGRIP FLO-41 delivery system while following the human readable expiry date printed on the primary and secondary package labeling, which are also listed in the table above. 2. If you received this communication directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by e-mailing it to [email protected], even if you do not have any inventory. Returning the customer reply form promptly will confirm your receipt of this notification and prevent you from receiving repeat notices. If you do not return the customer reply form, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 3. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them.
- Quantity in commerce
- 4,752 units
- Distribution
- United States, Australia, Hong Kong, New Zealand, Belgium, Cyprus, Ireland, Italy, Netherlands, Norway, Portugal, Spain, Sweden, and United Kingdom.
- Product codes and lots
- Product Code: ADS201897. Lot Numbers: 0000266641, 0000325839, 0000362640, 0000380266.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2051-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.