Z-2093-2021Class II

SoftGenomics version 4.1.15.6 SoftGenomics is a laboratory workflow and information management system to be...

SCC Soft Computer / initiated 2021-03-31 / Terminated / root cause: software

Soft Computer Consultants, Inc. began a recall of SoftGenomics version 4.1.15.6 SoftGenomics is a laboratory workflow and information management system to be used in a clinical molecular or cytogenetics... on . FDA classified it as Class II and gives the reason as "Software showing wrong results reported, PDF not match HIS". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2093-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SCC Soft Computer
Product
SoftGenomics version 4.1.15.6 SoftGenomics is a laboratory workflow and information management system to be used in a clinical molecular or cytogenetics laboratory or medical research laboratory setting by knowledgeable and trained users. SoftGenomics is used to record and track activities performed in ordering, specimen processing (pre-analytical steps), testing, documentation, and reporting.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Software showing wrong results reported, PDF not match HIS.
Root cause tag
software
FDA root cause
Software Design Change
Action
The firm sent an " URGENT: MEDICAL DEVICE RECALL" letter dated 6/2/2021 to its sole consignee of the correction on 06/02/2021 by mail. The letter described the product, problem and actions to be taken. The consignee was instructed to do the following: Acknowledge receipt of this letter in the TMS task Give permission to load Hot Fix on Test environment Perform validation activities to completion Acknowledge Hot Fix has been moved to Live environment and is working as expected Type of action by Company: All clients on affected lines of software have been notified of the recall. SCC will provide Hot Fixes for all affected lines that will be delivered to all affected clients. Until the software correction is delivered, clients have been informed of an alternative workflow to avoid the problem. The firm provided information on a software update performed on 03/31/2021 and stated that, "Hot Fix has already been delivered." If you have any further questions, Please call 1-800-763-8522 (Technical Support 24/7).
Quantity in commerce
1
Distribution
US Distribution to: CA only
Product codes and lots
Version 4.1.15.6

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2093-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2093-2021Class IITerminatedVoluntary: Firm initiated