Z-2013-2021Class II
Randox RX Imola Analyser with ISE, Catalog Numbers RX4900, RX4900R
Randox Laboratories, Limited / initiated 2021-04-28 / Terminated / root cause: software
Randox Laboratories Ltd. began a recall of Randox RX Imola Analyser with ISE, Catalog Numbers RX4900, RX4900R on . FDA classified it as Class II and gives the reason as "An issue was identified where the software froze during processing of commands, which resulted in no results displayed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2013-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Randox RX Imola Analyser with ISE, Catalog Numbers RX4900, RX4900R
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- An issue was identified where the software froze during processing of commands, which resulted in no results displayed.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Medical Device Correction letters dated 4/28/21 were sent to customers. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. " Please contact your local service provider to arrange for the software update to be performed . "Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. Please accept our apologies for any inconvenience caused. Thank you for your patience and understanding. If you have any questions or concerns please contact Randox Technical Services. Local Contact Information USA Randox Laboratories-US, Ltd. 515 Industrial Boulevard Kearneysville West Virginia, 25430 Tel: +1 304 728 2890 Toll Free 866 4 RANDOX Email: [email protected] Puerto Rico Clinical Diagnostics of P.R. LLC Carr. 887km 0.6, Carolina Commercial Park Local B-2 Carolina, Puerto Rico Tel: 1-787-701-7000 Email: [email protected]
- Quantity in commerce
- 50 kits
- Distribution
- US Nationwide Distribution and Puerto Rico.
- Product codes and lots
- GTIN 05055273206104, Serial Numbers: 7201-0423 7201-0417 7201-0353 7201-0776 7201-0367 7201-0921 7201-0982 7201-0977 7201-0970 7201-0950 7201-0924 7201-0919 7201-0901 7201-0849 7201-0827 7201-0799 7201-0801 7201-0768 7201-0606 7201-0562 7201-0535 7201-0531 7201-0381 7201-0112 7201-0231 7201-0275 7201-0299 7201-0319 7201-0343 7201-0348 7201-0391 7201-0514 7201-0515 7201-0540 7201-0605 7201-0616 7201-0840 7201-0533 7201-0932 7201-0943 7201-0513 7201-0482 7201-0960 7201-0882 7201-0434 7201-0607 7201-0768 7201-0961 7201-0980 7201-0994
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2013-2021 | Class II | Terminated | Voluntary: Firm initiated |