Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2141-2021Class II

OxyMask O2 Adult REF OM-1125-14

Southmedic Inc. / initiated 2021-05-25 / Open, Classified / root cause: manufacturing process

Southmedic, Inc. began a recall of OxyMask O2 Adult REF OM-1125-14 on . FDA classified it as Class II and gives the reason as "Potential for oxygen tubing to disconnect from mask or loosen. Disconnected tubing results in a loss of device function and use of the device could result in inappropriate or a lack of oxygen delivery to the patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2141-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Southmedic Inc.
Product
OxyMask O2 Adult REF OM-1125-14
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Potential for oxygen tubing to disconnect from mask or loosen. Disconnected tubing results in a loss of device function and use of the device could result in inappropriate or a lack of oxygen delivery to the patient.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 05/25/2021, the firm sent an "Urgent Medical Device Recall" Notice Letter via email with a follow-up telephone conversation on 06/03/2021 to inform customer that the Recalling Firm has become aware of the potential for disconnection of the oxygen tubing at the mask's elbow. This disconnection may result in a loss of device function leading to a lack of oxygen delivery to the patient. Customers are instructed to take the following actions: 1. Review your inventory and confirm whether you are still in possession of the impacted product. 2. Complete the Confirmation Form provided and fax the form to the attention of Tish Whitehead at 705-728-9537. 3. Please return any of the impacted products you may have in stock to the following address: Southmedic Inc., 364 St Vincent Street, Unit #15 Barrie, ON L4M 4A5 4. Please write RGA# OM112514MAY2021 on the outside of the package. Product may be returned using UPS Account # 5652381. For any questions - contact the VP Quality and Regulatory Affairs at (800) 463-7146 ext 342 or by email at [email protected].
Quantity in commerce
160 cases (25 units per case) in U.S.
Distribution
U.S.: IL O.U.S.: N/A
Product codes and lots
Model: OM-1125-14 Lot Number: W73887

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2141-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2141-2021Class IIOngoingVoluntary: Firm initiated