Z-1995-2021Class II

MaxPlusTM bi fuse extension set with 2 clear needleless connectors -

CareFusion 303, Inc / initiated 2021-06-09 / Terminated

CAREFUSION began a recall of MaxPlusTM bi fuse extension set with 2 clear needleless connectors - Product Usage: The needleless connector can be used for direct injection, intermittent... on . FDA classified it as Class II and gives the reason as "A portion of a validation lot was inadvertently released to the US Market. The product did not meet the validation testing criteria regarding leakage". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1995-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
MaxPlusTM bi fuse extension set with 2 clear needleless connectors - Product Usage: The needleless connector can be used for direct injection, intermittent infusion, continuous infusion, or aspiration.
Product code
FPA Set, Administration, Intravascular
Reason for recall
A portion of a validation lot was inadvertently released to the US Market. The product did not meet the validation testing criteria regarding leakage.
FDA root cause
Under Investigation by firm
Action
On 06/09/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via email and FedEx to customers informing them that a validation lot was inadvertently released to the US market. This product did not meet the validation testing criteria regarding leakage. Potential health consequences associated with the use of the non-validated product is under-infusion. Customers are instructed to: 1. Immediately review their inventory for the specific Catalog (Ref) and lot numbers listed. Destroy all product subject to the recall following their institutions process for destruction. The Recalling Firm will replace discarded product. 2. Share this recall notification with all users within or outside their facility network to which you they have further distributed the affected products to ensure they are also aware of this recall. 3. Complete the attached Customer Response Form and return to the Recalling Firm's contact noted on the form (BD Post Market Quality Email: [email protected], Or Fax No.: 312-949-0236) whether or not they have any of the impacted material so that the Recalling Firm may acknowledge their receipt of this notification per FDA requirements and subsequently process their product replacement. For any questions or assistance contact: North American Regional Complaint Center 1-84BD-LIFE (1-844-823-5433) Say Recall when prompted MonFri 8:00am and 5:00pm CT
Quantity in commerce
100 units
Distribution
U.S.: TN O.U.S.: None
Product codes and lots
Catalog Number:MP9246-C GTIN: 10885403237157 Lot Number: 20087194

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1995-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1995-2021Class IITerminatedVoluntary: Firm initiated