Z-2139-2021Class II

CARESCAPE PDM-Masimo SpO2. Physiological data monitor

GE Healthcare / initiated 2021-05-21 / Open, Classified / root cause: software

GE Healthcare, LLC began a recall of CARESCAPE PDM-Masimo SpO2. Physiological data monitor on . FDA classified it as Class II and gives the reason as "Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2139-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
CARESCAPE PDM-Masimo SpO2. Physiological data monitor.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
On May 21st the firm sent a letter to its customers with the following instructions: You can continue to use the CARESCAPE PDM with Masimo SET technology by completing the power down instructions below annually. 1. If the PDM is on a patient, provide alternate monitoring if needed, during the power down process. 2. Disconnect the PDM from the display device by removing the PDM from the mounting dock or by disconnecting the communication cable from the back of the PDM. 3. Open the battery door compartment and remove the battery (if present), for approximately 10 seconds. 4. Reinsert the battery and reconnect the PDM to the display device
Quantity in commerce
111,728 units
Distribution
worldwide distribution US nationwide, Albania, Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Fiji, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, State of Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Product codes and lots
CARESCAPE PDM: product codes SA3, SPX; model numbers 2042084-001, 2094504-001, 2016793-002, 2016793-003; Software kits, field replaceable units, and upgrade kits: part numbers 2034826-001, 2034826-002, 2034826-003, 2034826-004, 2034826-005, 2034826-006, 2034826-007, 2034826-008, 2034826-009, 2034826-010, 2034826-011, 2034826-012, 2034826-013, 2031069-010, 2045825-001, 2045825-002, 2045825-004, 2045825-005

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2139-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2139-2021Class IIOngoingVoluntary: Firm initiated