Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2108-2021Class III

3AA battery cradle (Part Number 045-001699-01) used with BeneVision TD60 Telepack (Model Number 115-047566-00...

Datascope Corp. / initiated 2021-05-27 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of 3AA battery cradle (Part Number 045-001699-01) used with BeneVision TD60 Telepack (Model Number 115-047566-00 /115-060044-00) and TM80 Telepack (Model Number... on . FDA classified it as Class III and gives the reason as "Telepack may not power on when using the 3AA battery cradle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2108-2021
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
3AA battery cradle (Part Number 045-001699-01) used with BeneVision TD60 Telepack (Model Number 115-047566-00 /115-060044-00) and TM80 Telepack (Model Number 115-047565-00 / 120-018867-00). Used with physiological data monitor.
Product code
MSX System, Network And Communication, Physiological Monitors
Reason for recall
Telepack may not power on when using the 3AA battery cradle.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Urgent Medical Device Correction notification letters dated 5/27/21 were sent to customers. Actions to be taken by the Customer/User Please discard all 3AA battery cradles in stock that were purchased between October 1, 2020 and March 31, 2021. Shipping records indicate your facility has XX cradles affected by this recall. Continue to use the TD60 and TM80 telepacks normally with either the 2AA battery cradle (TD60 only), or rechargeable Li-ion battery pack, if previously purchased. Please complete and return the Acknowledgement and Receipt form (found on the 3rd page of this recall notice) by June 30, 2021. Product and Distribution Information The 3AA cradles included with the TD60 or TM80 telepacks distributed between October 1, 2020 and March 31, 2021 may be affected by this recall. Other Information Please complete the attached Acknowledgement and Receipt Form and return in the stamped envelope provided. Contact information is provided below should you have questions. Adverse reactions or quality problems experienced with the use of this product may be reported to FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. We apologize for any inconvenience this may cause and appreciate your cooperation. Sincerely, Diane Arpino Director, Quality Operations and Regulatory Affairs Mindray DS USA, Inc. [email protected] 201.995.8407 Distributors were instructed to do the following: Actions to be taken by the Customer/User Please discard all 3AA battery cradles in stock that were purchased between October 1, 2020 and March 31, 2021. Shipping records indicate your facility purchased 6 cradles affected by this recall. Continue to use the TM80 telepack normally with rechargeable Li-ion battery pack, if previously purchased. Please contact each customer which you have distributed the affected 3AA battery cradles. The format for the recall notification to send to your customers is provided. We recommend that you send the notificati
Quantity in commerce
804
Distribution
Distributed US nationwide and Canada.
Product codes and lots
Part Number 045-001699-01;  Awaiting additional details

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2108-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2108-2021Class IIITerminatedVoluntary: Firm initiated