Z-2100-2021Class II
Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962
Howmedica Osteonics Corp / initiated 2021-05-18 / Terminated
Stryker Spine began a recall of Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962 on . FDA classified it as Class II and gives the reason as "The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2100-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962
- Product code
- NDN Cement, Bone, Vertebroplasty
- Reason for recall
- The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.
- FDA root cause
- Storage
- Action
- Urgent Medical Device Recall notification letters dated 5/14/21 were sent to customers. Actions needed NOTE: due to the potential hazards/risks associated with use of product associated with these shipments, please urgently quarantine the in-scope product. 1. Immediately review the Product Recall Notification. Please forward this information to all appropriate personnel within your organization. Maintain awareness of this communication internally until all required actions have been completed within your facility. 2. Immediately check your internal inventory to locate the affected products. Quarantine and discontinue use of any affected Cortoss Bone Augmentation Material. Your local Stryker Representative is available to assist with removal of this product. 3. Complete the enclosed business reply form (BRF) to confirm receipt of this Notification and identify how many affected items are currently in your inventory. Please return the BRF, even if you dont have any affected product on hand, via email to [email protected]. 4. Upon receipt of the completed BRF, replacement(s) will be shipped to your facility. 5. If you have further distributed this product to other organizations, please provide contact details on the BRF so that Stryker can inform the recipients appropriately. If you have any questions or concerns, please contact Regulatory Compliance at 201-749-8090.
- Quantity in commerce
- 559 total
- Distribution
- US Nationwide Distribution
- Product codes and lots
- lot A1908003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2100-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2101-2021 | Class II | Terminated | Voluntary: Firm initiated |