Z-2226-2021Class II
ETEST IMIPENEM RELEBACTAM. in vitro diagnostic
bioMerieux, Inc. / initiated 2021-06-24 / Terminated
Biomerieux Inc began a recall of ETEST IMIPENEM RELEBACTAM. in vitro diagnostic on . FDA classified it as Class II and gives the reason as "Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2226-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- ETEST IMIPENEM RELEBACTAM. in vitro diagnostic
- Product code
- JWY Manual Antimicrobial Susceptibility Test Systems
- Reason for recall
- Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains.
- FDA root cause
- Under Investigation by firm
- Action
- The recall notification was issued to the impacted subsidiaries and distributors on 24JUN2021. The notification includes a Customer Letter to be translated into local language and sent to each impacted customer. The customer letter will inform the customers about the MIC overestimation and false resistant results issue and will be instructed to: - Perform a retest using an alternative method on P. aeruginosa strains for which they obtained resistant or intermediate results with ETEST IMIPENEM RELEBACTAM ref. 420925 and 420927 - Among tests previously performed, identify any possible false resistant/intermediate results that may have occurred, to analyze the related risks and to determine appropriate actions if relevant.
- Quantity in commerce
- 1,431
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii. The countries of Austria, Belgium, Croatia, Czech Republic, France, Germany, Greece, Italy, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland and United Kingdom.
- Product codes and lots
- Ref 420927 (US Product) Lot numbers 1007888860, 1008598340, 1008701580, 1007607290, 1007888870, 1008390030, 1008598350 and 1008701590 and Ref 420925 (not distributed in US):Lot Codes: 1007888860, 100859840 and 1008701580
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2226-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY