Gore Propaten Vascular Graft configured for Pediatric Shunt, Diameter 5 mm, Length 10 cm, Rx Only, Sterile...
W L Gore & Associates, Inc. began a recall of Gore Propaten Vascular Graft configured for Pediatric Shunt, Diameter 5 mm, Length 10 cm, Rx Only, Sterile EO, for the following Reference numbers: HPT050010... on . FDA classified it as Class II and gives the reason as "Vascular Grafts containing an inner diameter of 6 mm may be incorrectly labeled as 5 mm and Vascular Grafts containing an inner diameter of 5 mm are incorrectly labeled as 6 mm". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2292-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- W. L. Gore & Associates, Inc.
- Product
- Gore Propaten Vascular Graft configured for Pediatric Shunt, Diameter 5 mm, Length 10 cm, Rx Only, Sterile EO, for the following Reference numbers: HPT050010 UDI: (01)00733132606320 HPT050010A UDI: (01)00733132606337
- Product code
- DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
- Reason for recall
- Vascular Grafts containing an inner diameter of 6 mm may be incorrectly labeled as 5 mm and Vascular Grafts containing an inner diameter of 5 mm are incorrectly labeled as 6 mm.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- On May 10, 2021, Gore issued a "Medical Device Recall" notification to affected customers via USPS. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. Identify and return any unused devices within the scope of this voluntary recall. 2. Please complete and sign the enclosed CUSTOMER RETURN FORM and return to [email protected] within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. Please transfer this notice to other organization(s) as appropriate. 4. Any adverse event involving the GORE PROPATEN Vascular Graft configured for Pediatric Shunt should be reported to the manufacturer and the country specific regulatory authorities immediately. A. To report an event to W. L. Gore & Associates, email: [email protected], or contact: USA: +1 800 528 1866 Ext. 44922 / +1 928 864 4922, Fax +1 928 864 4364 EMEA: +49 89 4612 3440, Fax +49 89 4612 43440 Brazil: +55 11 5502-7953, Fax: +55 11 5502-7965 B. For US Customers only: Healthcare professionals and consumers may report adverse events or quality problems directly to FDA using the FDA MedWatch Website:https://www.accessdata.fda.gov/scripts/medwatch/index.cfm 5. Please feel free to contact Gore Customer Service (email: [email protected]) or Uzma Tahir, Ph.D. (928-864-3779/[email protected]) if you have questions or concerns regarding this voluntary recall.
- Distribution
- US: NE, DC, MA, KY, IL, WA, MD, CA, AZ, GA OUS: Canada, Brazil, Europe, Saudi Arabia
- Product codes and lots
- Serial Numbers for HPT050010A: 6185494PP001, 6185494PP002, 6185494PP004, 6185494PP006, 6185494PP008, 6185494PP009, 6185494PP010, 6185494PP011, 6185494PP012, 6185495PP001, 6185495PP002, 6185495PP004, 6185495PP005, 6185495PP006, 6185495PP007, 6185495PP009, 6185495PP010, 6185495PP011, 6185495PP012, 6185495PP014 and Serial Numbers for HPT050010: 6185497PP001, 6185497PP005, 6185497PP006, 6185497PP007, 6185497PP008, 6185497PP009, 6185497PP010, 6185497PP011, 6185497PP012
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2292-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2293-2021 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY