Z-2165-2021Class I
Dose IQ Safety Software used with Spectrum IQ Infusion Pump
Baxter Healthcare Corporation / initiated 2021-07-07 / Terminated / root cause: software
Baxter Healthcare Corporation began a recall of Dose IQ Safety Software used with Spectrum IQ Infusion Pump on . FDA classified it as Class I and gives the reason as "Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2165-2021
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Dose IQ Safety Software used with Spectrum IQ Infusion Pump
- Product code
- FRN Pump, Infusion
- Reason for recall
- Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued an Urgent Medical Device Correction notice by letter on July 7, 2021. The letter explained the issue and provided a temporary workaround which can be implemented until the new software tool is ready for distribution.
- Quantity in commerce
- 61 units
- Distribution
- US Nationwide Distribution.
- Product codes and lots
- Software version: All libraries initially created with version 9.0.x
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2165-2021 | Class I | Terminated | Voluntary: Firm initiated |