Z-2165-2021Class I

Dose IQ Safety Software used with Spectrum IQ Infusion Pump

Baxter Healthcare Corporation / initiated 2021-07-07 / Terminated / root cause: software

Baxter Healthcare Corporation began a recall of Dose IQ Safety Software used with Spectrum IQ Infusion Pump on . FDA classified it as Class I and gives the reason as "Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2165-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Dose IQ Safety Software used with Spectrum IQ Infusion Pump
Product code
FRN Pump, Infusion
Reason for recall
Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
Root cause tag
software
FDA root cause
Software design
Action
The firm issued an Urgent Medical Device Correction notice by letter on July 7, 2021. The letter explained the issue and provided a temporary workaround which can be implemented until the new software tool is ready for distribution.
Quantity in commerce
61 units
Distribution
US Nationwide Distribution.
Product codes and lots
Software version: All libraries initially created with version 9.0.x

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2165-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2165-2021Class ITerminatedVoluntary: Firm initiated