Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to...
Cardinal Health 200, LLC began a recall of Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the... on . FDA classified it as Class I and gives the reason as "The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2244-2021
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201
- Product code
- FOS Catheter, Umbilical Artery
- Reason for recall
- The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Cardinal Health issued Urgent Medical Device Correction letter and Acknowledgement Form on June 16, 2021 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1) REVIEW the IFU for the Safety Scalpel N11 (vendor part number 73-1811). 2) COMMUNICATE the IFU to all personnel that utilize the Argyle" UVC Insertion Tray (SKU 43201) containing Safety Scalpel N11 (vendor part number 73-1811) 3) NOTIFY any customers to whom you may have distributed/forwarded affected product or will send the product about this product correction notice and share a copy of this notice and the attached IFU. 4) POST a copy of this notification in the location where the product is stored. 5) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected] whether or not you have affected product. Available Assistance: Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All other customers 888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
- Quantity in commerce
- 7197 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- All lot codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2244-2021 | Class I | Terminated | Voluntary: Firm initiated |