Z-2237-2021Class II
Advance Dx 100 Blood Collection Card Quantity 25 REF CMT01 Advance Dx 100 Blood Collection Card and Foil Bags...
Advance Dx, Inc. / initiated 2021-07-06 / Terminated / root cause: manufacturing process
Advance Dx, Inc. began a recall of Advance Dx 100 Blood Collection Card Quantity 25 REF CMT01 Advance Dx 100 Blood Collection Card and Foil Bags Quantity 500 Cards and Foil Bags REF CMT01/F02... on . FDA classified it as Class II and gives the reason as "Due to high glucose test results when using the blood collection cards". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2237-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advance Dx, Inc.
- Product
- Advance Dx 100 Blood Collection Card Quantity 25 REF CMT01 Advance Dx 100 Blood Collection Card and Foil Bags Quantity 500 Cards and Foil Bags REF CMT01/F02 IFU: Advance Dx 100 SSE(Serum Separating Extraction) Technology BLOOD COLLECTION INSTRUCTIONS
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- Due to high glucose test results when using the blood collection cards.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 06/15/2021, the firm sent an initial customer notification via email and on 07/06/2021 follow up email was sent with a revised "FOLLOW-UP URGENT: MEDICAL DEVICE RECALL" Letter. The notification was to inform customers that blood collection cards manufactured in 2020 provided inaccurate blood glucose values. Customers are instructed to: 1. If they are conducting blood glucose testing with the affected Blood Collection Card, immediately STOP and review their inventory and remove them from all inventory locations. Recalled products can be identified by the Lot number printed on the products packaging and shipper containers. 2. Pass on this information to anyone who uses or orders Blood Collection Cards at their facility. Additionally, ensure that a copy of this letter is provided to any other organizations to which affected devices have been transferred. 3. Complete the Customer Recall Response Card, scan and email immediately to the email address on the Customer Recall Response Card for product in your facility at the time of receipt of this letter. 4. The Recalling Firm is working to ensure that replacement products are readily available. Once customers have identified how much product they will be returning, call Recalling Firm's Customer Service at 1-847-677-3823 to arrange for product replacement. Any questions concerning this product recall, contact Director of Regulatory Affairs/Quality Assurance at 1-949-433-3058 between 8:00 AM and 5:00 PM Pacific Standard Time, Monday-Friday.
- Quantity in commerce
- 179,750 cards
- Distribution
- U.S. Nationwide distribution in the states of MD and NC.
- Product codes and lots
- Product Catalog Number: CMT01 GTIN: 1061C010890976 Lot Number: PPD02102 V6
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2237-2021 | Class II | Terminated | Voluntary: Firm initiated |