Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2262-2021Class II

Arcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 50 MM Item Number: 11-301310

Zimmer GmbH / initiated 2021-06-02 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Arcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 50 MM Item Number: 11-301310 on . FDA classified it as Class II and gives the reason as "Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2262-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Arcos Modular Revision Hip System High Offset Cone Prox Body, Size A, 50 MM Item Number: 11-301310
Product code
KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Reason for recall
Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer Biomet issued Urgent Medical Device Notification Letter on 6/2/21 to Distributors/ Risk Managers/Surgeons, stated reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please ca
Quantity in commerce
81 units
Distribution
US Nationwide distribution.
Product codes and lots
Lot Numbers/UDI : 004230 (01)00880304473799(17)310412(10)004230 626590 (01)00880304473799(17)310312(10)626590 626600 (01)00880304473799(17)310314(10)626600 678840 (01)00880304473799(17)310305(10)678840 679170 (01)00880304473799(17)310423(10)679170 679180 (01)00880304473799(17)310318(10)679180 880170 (01)00880304473799(17)310409(10)880170

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2262-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2259-2021Class IITerminatedVoluntary: Firm initiated
Z-2260-2021Class IITerminatedVoluntary: Firm initiated
Z-2261-2021Class IITerminatedVoluntary: Firm initiated
Z-2262-2021Class IITerminatedVoluntary: Firm initiated
Z-2263-2021Class IITerminatedVoluntary: Firm initiated
Z-2264-2021Class IITerminatedVoluntary: Firm initiated
Z-2265-2021Class IITerminatedVoluntary: Firm initiated
Z-2266-2021Class IITerminatedVoluntary: Firm initiated
Z-2267-2021Class IITerminatedVoluntary: Firm initiated
Z-2268-2021Class IITerminatedVoluntary: Firm initiated
Z-2269-2021Class IITerminatedVoluntary: Firm initiated
Z-2270-2021Class IITerminatedVoluntary: Firm initiated
Z-2271-2021Class IITerminatedVoluntary: Firm initiated
Z-2272-2021Class IITerminatedVoluntary: Firm initiated
Z-2273-2021Class IITerminatedVoluntary: Firm initiated
Z-2274-2021Class IITerminatedVoluntary: Firm initiated
Z-2275-2021Class IITerminatedVoluntary: Firm initiated