Z-2269-2021Class II
Arcos Modular Revision Hip System Standard Cone Prox Body, Size E, 70 MM Item Number:11-301325
Zimmer GmbH / initiated 2021-06-02 / Terminated / root cause: manufacturing process
Biomet, Inc. began a recall of Arcos Modular Revision Hip System Standard Cone Prox Body, Size E, 70 MM Item Number:11-301325 on . FDA classified it as Class II and gives the reason as "Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2269-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Arcos Modular Revision Hip System Standard Cone Prox Body, Size E, 70 MM Item Number:11-301325
- Product code
- KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
- Reason for recall
- Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Zimmer Biomet issued Urgent Medical Device Notification Letter on 6/2/21 to Distributors/ Risk Managers/Surgeons, stated reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please ca
- Quantity in commerce
- 11 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot Numbers/UDI : 880850 (01)00880304480575(17)310412(10)880850
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2259-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2260-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2261-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2262-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2263-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2264-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2265-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2266-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2267-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2268-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2269-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2270-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2271-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2272-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2273-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2274-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2275-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.