Z-2238-2021Class II
Prelude SNAP Splittable Sheath Introducer, REF: PLSX -1010, size 10F, Standard Length: 13 cm, CE0344, STERILE...
Merit Medical Systems, Inc. / initiated 2021-06-30 / Terminated / root cause: packaging
Merit Medical Systems, Inc. began a recall of Prelude SNAP Splittable Sheath Introducer, REF: PLSX -1010, size 10F, Standard Length: 13 cm, CE0344, STERILE EO, Rx Only, (01) 00884450489611. For the... on . FDA classified it as Class II and gives the reason as "There is a potential that 7F sheaths are packaged as 10F sheaths". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2238-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Prelude SNAP Splittable Sheath Introducer, REF: PLSX -1010, size 10F, Standard Length: 13 cm, CE0344, STERILE EO, Rx Only, (01) 00884450489611. For the introduction of various types of pacing leads and catheters to the heart and coronary venous system.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- There is a potential that 7F sheaths are packaged as 10F sheaths.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- On June 30, 2021, Merit Medical issued an "Urgent Medical Device Recall Notice" to affected consignees via FedEx overnight. In addition, to informing consignees about the recall, the firm ask consignees to take the following action: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed Customer Response Form to Merit Customer Service at [email protected] within 10 business days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. 6. Adverse events experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online at www.fda.gov/medwatch or by phone at 1-800-332-1088. 7. If you have any questions concerning this communication, please dont hesitate to contact your Merit Sales Representative or Merit Customer Service at (801) 208-4381 | Hours: 6 am to 6 pm MDT | Mon-Fri.
- Quantity in commerce
- 760 units
- Distribution
- US:AL,AR,CA,CO,DE,FL,GA,HI,IA,IL,KY,MA,MD,ME,MI,MN,NC,NE,NJ,NV,NY,OH,OR,PA,RI,TX,VA,VT, WI OUS: Norway
- Product codes and lots
- Lot #I2067768
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2238-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2239-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY