Z-2312-2021Class II
NEO-KATH Epidural Catheter Set, REF 200-2430, Sterile
Epimed International, Inc. / initiated 2017-12-28 / Completed / root cause: manufacturing process
Epimed International began a recall of NEO-KATH Epidural Catheter Set, REF 200-2430, Sterile on . FDA classified it as Class II and gives the reason as "Potential for stylet to protrude from the distal end of the catheter". As of , the recall status on record is Completed, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2312-2021
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Epimed International, Inc.
- Product
- NEO-KATH Epidural Catheter Set, REF 200-2430, Sterile
- Product code
- BSO Catheter, Conduction, Anesthetic
- Reason for recall
- Potential for stylet to protrude from the distal end of the catheter
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On about 12/28/2017, Epimed mailed their recall notification letter titled, URGENT Medical Device RECALL, to customers. Those letters were followed up with a telephone call or email containing the letter on about 12/30/2017. The letter explained the issue and hazard and requested that if any potentially affected product remains in the accounts inventory, Epimed will offer to replace/credit the affected product. Each customer was asked to complete and return the RECALL RETURN RESPONSE FORM as soon as possible.
- Quantity in commerce
- 132 units
- Distribution
- Worldwide distribution, with US distribution to CA, DE, OH, OK, PA, VA, WA, WV. International distribution to Canada and Iran.
- Product codes and lots
- Reference Number 200-2430, Lot Numbers 16308275 (Expiration August 2022), 16308382 (Expiration October 2022)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2312-2021 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-2313-2021 | Class II | Completed | Voluntary: Firm initiated |