Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system
Philips North America Llc began a recall of Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2315-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Model 728321/728323 CT5000 Ingenuity Plus/Pro/Premium-Computed Tomography X-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to [email protected] The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 2 units
- Distribution
- US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
- Product codes and lots
- Model: 728321 S/N UDI: 40012 (01)00884838059498(21)40012 Model: 728332 40004 (01)00884838059504(21)40004 Software Version: 4.1.10 SynchRight Option
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2314-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2315-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2316-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2317-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2318-2021 | Class II | Terminated | Voluntary: Firm initiated |