Z-2316-2021Class II

728327 Ingenuity CT Upgrades-Computed Tomography X-ray system

Philips North America LLC / initiated 2021-06-08 / Terminated / root cause: software

Philips North America Llc began a recall of 728327 Ingenuity CT Upgrades-Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2316-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
728327 Ingenuity CT Upgrades-Computed Tomography X-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Root cause tag
software
FDA root cause
Software design
Action
Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to [email protected] The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Quantity in commerce
3 units
Distribution
US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
Product codes and lots
Serial Number: 300099 300106 300143 Software Version 4.1.10 SynchRight Option

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2316-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2314-2021Class IITerminatedVoluntary: Firm initiated
Z-2315-2021Class IITerminatedVoluntary: Firm initiated
Z-2316-2021Class IITerminatedVoluntary: Firm initiated
Z-2317-2021Class IITerminatedVoluntary: Firm initiated
Z-2318-2021Class IITerminatedVoluntary: Firm initiated