Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for...
Baxter Healthcare Corporation began a recall of Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement... on . FDA classified it as Class II and gives the reason as "Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-2379-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- The firm notified its consignees of the recall by letter on 07/22/2021. The letter explained the issue and requested the return of the affected product. Distributors were directed to notified their consignees.
- Quantity in commerce
- 1976 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot Numbers: 20B2330M
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2378-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2379-2021 | Class II | Terminated | Voluntary: Firm initiated |