Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598A
Cordis Corporation began a recall of Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598A on . FDA classified it as Class I and gives the reason as "Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2430-2021
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598A
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated an Urgent Medical Device Field Correction notice beginning on 07/21/2021. The letter described the problem and provided recommendations for clinical use found in the instructions for use.
- Quantity in commerce
- 18730 units
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.
- Product codes and lots
- all codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2428-2021 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2429-2021 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2430-2021 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2431-2021 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2432-2021 | Class I | Terminated | Voluntary: Firm initiated |