Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2431-2021Class I

Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598B

Cordis US Corp. / initiated 2021-07-21 / Terminated / root cause: design

Cordis Corporation began a recall of Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598B on . FDA classified it as Class I and gives the reason as "Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2431-2021
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598B
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.
Root cause tag
design
FDA root cause
Device Design
Action
The firm disseminated an Urgent Medical Device Field Correction notice beginning on 07/21/2021. The letter described the problem and provided recommendations for clinical use found in the instructions for use.
Quantity in commerce
71330 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.
Product codes and lots
all codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2431-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2428-2021Class ITerminatedVoluntary: Firm initiated
Z-2429-2021Class ITerminatedVoluntary: Firm initiated
Z-2430-2021Class ITerminatedVoluntary: Firm initiated
Z-2431-2021Class ITerminatedVoluntary: Firm initiated
Z-2432-2021Class ITerminatedVoluntary: Firm initiated