Z-2471-2021Class II
ACUSON Juniper Diagnostic Ultrasound System
Siemens Medical Solutions USA, Inc. / initiated 2021-07-12 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc. began a recall of ACUSON Juniper Diagnostic Ultrasound System on . FDA classified it as Class II and gives the reason as "Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2471-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ACUSON Juniper Diagnostic Ultrasound System
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will be contacted by a Customer Service Engineer to schedule a visit to their facility to update the system. Customers are asked to ensure that all users are aware of this issue and correction within their organization and others who may need to be inform of this situation. For further questions, contact Siemens Healthineers USD Complaints at [email protected]
- Quantity in commerce
- 15 systems
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.
- Product codes and lots
- Model Number: 11335791 UDI: (01)04056869152059 (21)123456 (240)11335791 (422)410 Firmware version 3.8 for rev. 7 U.S. Serial Number: 555606; O.U.S. Serial Numbers: 550975; 552770; 553629; 553951; 555060; 555601; 555602; 555604; 555605; 555610; 555611; 555612; 555613; 555626; O.U.S. Inactive Serial Numbers: 555603; 555607;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2471-2021 | Class II | Terminated | Voluntary: Firm initiated |