Z-2471-2021Class II

ACUSON Juniper Diagnostic Ultrasound System

Siemens Medical Solutions USA, Inc. / initiated 2021-07-12 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of ACUSON Juniper Diagnostic Ultrasound System on . FDA classified it as Class II and gives the reason as "Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2471-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON Juniper Diagnostic Ultrasound System
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.
Root cause tag
software
FDA root cause
Software change control
Action
On 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will be contacted by a Customer Service Engineer to schedule a visit to their facility to update the system. Customers are asked to ensure that all users are aware of this issue and correction within their organization and others who may need to be inform of this situation. For further questions, contact Siemens Healthineers USD Complaints at [email protected]
Quantity in commerce
15 systems
Distribution
Worldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.
Product codes and lots
Model Number: 11335791 UDI: (01)04056869152059 (21)123456 (240)11335791 (422)410 Firmware version 3.8 for rev. 7 U.S. Serial Number: 555606; O.U.S. Serial Numbers: 550975; 552770; 553629; 553951; 555060; 555601; 555602; 555604; 555605; 555610; 555611; 555612; 555613; 555626; O.U.S. Inactive Serial Numbers: 555603; 555607;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2471-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2471-2021Class IITerminatedVoluntary: Firm initiated