FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No...

BioFire Diagnostics, LLC / initiated 2021-07-20 / Terminated / root cause: manufacturing process

BioFire Diagnostics, LLC began a recall of FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127 on . FDA classified it as Class II and gives the reason as "Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2245-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioFire Diagnostics, LLC
Product
FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127
Product code
PEN Gram-Negative Bacteria And Associated Resistance Markers
Reason for recall
Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 07/20/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customer informing them that an issue has been identified in the manufacturing process which might contribute to elevated rates of false negative results while using the BCID Panel. Actions to be taken by customers: 1) Immediately examine their inventory for product identified in this recall, 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3) Complete the accompanying Field Action Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to the Recalling Firm so that it may acknowledge their receipt of this notification. 4) If they may have further distributed the affected product, identify any recipients and notify them at once of this product recall. For product replacement, as well as any other questions or concerns, contact customer support department at [email protected] or via telephone at 1-800-735-6544 and selecting option 5 for Product Technical Support.
Quantity in commerce
29 kits
Distribution
U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None
Product codes and lots
Part No: RFIT-ASY-0126 / UDI: 00815381020086;  Part No: RFIT-ASY-0127 / UDI: 00815381020093;  Kit Lot# 0868221; Pouch Lot# 194Z21; Batch 003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2245-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2245-2021Class IITerminatedVoluntary: Firm initiated