Z-2443-2021Class II

The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app

HERO HEALTH, Inc. / initiated 2021-07-02 / Terminated

Hero Health began a recall of The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app on . FDA classified it as Class II and gives the reason as "Due to dispenser prompting for a missed dose that had already been dispensed". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2443-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HERO HEALTH, Inc.
Product
The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app.
Product code
NXB Dispenser, Solid Medication
Reason for recall
Due to dispenser prompting for a missed dose that had already been dispensed
FDA root cause
Under Investigation by firm
Action
On 07/02/2021, the firm initially sent out an "URGENT NOTICE" via email notifying customers of an product update correction. Customers were informed that the firm had become aware that a recent firmware update (v2.4.12) that users have experienced an error, which caused improper recording of a dispensed dose as having not been dispensed. This error may prompt a user to dispense the same dose again, and the mobile app will indicate that the previously dispensed dose has not been taken yet. Customers are instructed to: -If the pill dispenser prompts you to re-dispense a scheduled dose which has already been dispensed today, select Skip. -Use extra caution when dispensing doses prior to the scheduled firmware update and verify that the pill dispenser has a working internet connection. This will ensure that the firmware update (v2.4.14) will be automatically applied on all potentially affected online pill dispensers (at approximately 2:00 AM on July 3, 2021) to correct the issue. -To manually update firmware immediately, follow these steps: 1) Ensure the Pill Dispenser is connected to WiFi (Check for the WiFi icon on the main screen). 2) Go to the Device menu, by pressing the Center button on the main screen. 3) Scroll down and select Device Settings . 4) Scroll down and select Firmware Update. 5) Select I m Ready. The update could take several minutes. It is recommended to run the firmware update at least 30 minutes before a scheduled dose. -If the pill dispenser is not connected to WiFi, the firmware update cannot be completed. In this event, contact the Recalling Firm's Member support for assistance before continuing to use the pill dispenser to ensure the update is applied as soon as possible For question or further assistance, contact the Recalling firm at 855-855-9962 or via email at [email protected] (24/7).
Quantity in commerce
12,039 units which had software version V2.04.012
Distribution
U.S. Nationwide distribution.
Product codes and lots
Model 100 (H-100) Catalog Number: F00002 Hero 100 V1.4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2443-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2443-2021Class IITerminatedVoluntary: Firm initiated