Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2285-2021Class I

Monoject 0.9% Sodium Chloride Flush Syringe, 3 mL Fill, STERILE, Product Code 8881570123 The proposed device...

Medtronic, Inc. / initiated 2021-08-04 / Open, Classified / root cause: design

Cardinal Health began a recall of Monoject 0.9% Sodium Chloride Flush Syringe, 3 mL Fill, STERILE, Product Code 8881570123 The proposed device is indicated for use in flushing compatible... on . FDA classified it as Class I and gives the reason as "Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2285-2021
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Monoject 0.9% Sodium Chloride Flush Syringe, 3 mL Fill, STERILE, Product Code 8881570123 The proposed device is indicated for use in flushing compatible intravenous tubing systems and indwelling intravascular access devices.
Product code
NGT Saline, Vascular Access Flush
Reason for recall
Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe.
Root cause tag
design
FDA root cause
Device Design
Action
On August 4, 2021 Cardinal Health issued URGENT MEDICAL DEVICE RECALL notices and response forms to customers via overnight delivery. Customers are advised to take the following actions: 1) REVIEW your inventory for affected product. 2) SEGREGATE and QUARANTINE all on-hand product per the Affected Product table. 3) COMMUNICATE the notice to clinical staff through safety huddles to ensure awareness of this product issue. 4) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether or not you have affected product. 5) CONTACT the appropriate Customer Service group to arrange for return and credit of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440 6) CUSTOMERS that did not receive product directly from Cardinal Health should contact the location where they purchased it. For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
Quantity in commerce
267,217,860 eaches in total
Distribution
US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
Product codes and lots
20I1524 20D0694 20I1584

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2285-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2284-2021Class IOngoingVoluntary: Firm initiated
Z-2285-2021Class IOngoingVoluntary: Firm initiated
Z-2286-2021Class IOngoingVoluntary: Firm initiated