Monoject 0.9% Sodium Chloride Flush Syringe, 5mL Fill, STERILE, Product Code 8881570125 The proposed device...
Cardinal Health began a recall of Monoject 0.9% Sodium Chloride Flush Syringe, 5mL Fill, STERILE, Product Code 8881570125 The proposed device is indicated for use in flushing compatible... on . FDA classified it as Class I and gives the reason as "Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2286-2021
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Monoject 0.9% Sodium Chloride Flush Syringe, 5mL Fill, STERILE, Product Code 8881570125 The proposed device is indicated for use in flushing compatible intravenous tubing systems and indwelling intravascular access devices.
- Product code
- NGT Saline, Vascular Access Flush
- Reason for recall
- Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On August 4, 2021 Cardinal Health issued URGENT MEDICAL DEVICE RECALL notices and response forms to customers via overnight delivery. Customers are advised to take the following actions: 1) REVIEW your inventory for affected product. 2) SEGREGATE and QUARANTINE all on-hand product per the Affected Product table. 3) COMMUNICATE the notice to clinical staff through safety huddles to ensure awareness of this product issue. 4) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether or not you have affected product. 5) CONTACT the appropriate Customer Service group to arrange for return and credit of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440 6) CUSTOMERS that did not receive product directly from Cardinal Health should contact the location where they purchased it. For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
- Quantity in commerce
- 267,217,860 eaches in total
- Distribution
- US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
- Product codes and lots
- 20D0614 21D0894 20D0624 20D0674 20D0684 20H1504 20H1514
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2284-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2285-2021 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2286-2021 | Class I | Ongoing | Voluntary: Firm initiated |