Z-2438-2021Class II
Solero Generator Refurbished PG, Item No. H78712740000D0, Catalog No. 12740000D
Boston Scientific Corporation / initiated 2021-07-20 / Terminated / root cause: software
Angiodynamics, Inc. began a recall of Solero Generator Refurbished PG, Item No. H78712740000D0, Catalog No. 12740000D on . FDA classified it as Class II and gives the reason as "Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2438-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Solero Generator Refurbished PG, Item No. H78712740000D0, Catalog No. 12740000D
- Product code
- NEY System, Ablation, Microwave And Accessories
- Reason for recall
- Specific serial numbers of the Solero MTA Generator require servicing to upgrade the software to help reduce the incidence of Error 0001, which can occur during system start-up.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On July 20, 2021, the firm initiated the recall by issuing Urgent Voluntary Medical Device Recall letters to affected consignees. Customers were informed of the software upgrade. Customers were asked to return their generators to AngioDynamics for service. The firm's Hardware Service Department will contact you to schedule the upgrade, which may include the shipment of a replacement generator to your facility if needed to be used during the upgrade of your devices. In the interim, while this upgrade is being scheduled, you may continue to use the Solero generator and are instructed to adhere to the following Warning in the current product labeling (User Manual): WARNING: Do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "Ready".
- Quantity in commerce
- 8 OUS
- Distribution
- Domestic distribution Nationwide. Foreign distribution worldwide.
- Product codes and lots
- Serial Nos. QBY0002816 QBY0002817 QBY0002348 QBY0002308 QBY0002311 QBY0002711 QBY0002741 QBY0002858
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2436-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2437-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2438-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2439-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.