Z-2406-2021Class II

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Elekta Limited / initiated 2021-07-26 / Open, Classified

Elekta Inc began a recall of Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10 on . FDA classified it as Class II and gives the reason as "If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2406-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Elekta Limited
Product
Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
FDA root cause
Under Investigation by firm
Action
Elekta will send an Important Field Safety Notice 382-01-MON-019 to all affected customers beginning August 10, 2021. The notice informs users of the specific product and version numbers affected by the issue, and any workaround that can be used to avoid the issue. A copy of the Important Field Safety Notice is to be kept with the most current labeling and all personnel working with the product should be made aware of the content of the letter. Customers are asked to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible, but no later than 30 days after receipt. This issue is expected to be addressed in a future Monaco release, tentatively scheduled for the end of June 2022.
Distribution
Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom
Product codes and lots
Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2406-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2406-2021Class IIOngoingVoluntary: Firm initiated