LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402...

Lymol Medical Corporation / initiated 2021-07-29 / Terminated

Lymol Medical Corporation began a recall of LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210... on . FDA classified it as Class II and gives the reason as "Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2562-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lymol Medical Corporation
Product
LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420
Product code
EOQ Bronchoscope (Flexible Or Rigid)
Reason for recall
Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.
FDA root cause
Under Investigation by firm
Action
On August 9, 2021, the firm began notifying all consignees via email that the reprocessing instructions have changed. Included with each notification is a copy of the Urgent Notice, the revised IFU, and a response form. Customers were advised that the sterilization procedures were updated on November 15, 2020. Customers should discard and replace the obsolete procedures with the revised procedures included with the firm's mailing. The revised instructions have been uploaded to the LYMOL website at https://www.lymolmedical.com/instructions.cfm Customers with questions may contact the firm at [email protected].
Quantity in commerce
309
Distribution
US Nationwide distribution.
Product codes and lots
All distributed lots