LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402...
Lymol Medical Corporation began a recall of LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210... on . FDA classified it as Class II and gives the reason as "Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2562-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lymol Medical Corporation
- Product
- LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.
- FDA root cause
- Under Investigation by firm
- Action
- On August 9, 2021, the firm began notifying all consignees via email that the reprocessing instructions have changed. Included with each notification is a copy of the Urgent Notice, the revised IFU, and a response form. Customers were advised that the sterilization procedures were updated on November 15, 2020. Customers should discard and replace the obsolete procedures with the revised procedures included with the firm's mailing. The revised instructions have been uploaded to the LYMOL website at https://www.lymolmedical.com/instructions.cfm Customers with questions may contact the firm at [email protected].
- Quantity in commerce
- 309
- Distribution
- US Nationwide distribution.
- Product codes and lots
- All distributed lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2562-2021 | Class II | Terminated | Voluntary: Firm initiated |