Z-2502-2021Class II

Centricity PACS-IW with Universal Viewer -

GE Healthcare / initiated 2021-08-06 / Open, Classified / root cause: software

GE Healthcare, LLC began a recall of Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources on . FDA classified it as Class II and gives the reason as "Image acquisition failures and synchronization failure with the Centricity Enterprise Archive". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2502-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
Root cause tag
software
FDA root cause
Software design
Action
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
392 (219 US; 173 OUS) all products
Distribution
AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.
Product codes and lots
2068177-001 (Centricity PACS-IW with Universal Viewer software versions 5.0.x with PACS-IW foundation) System ID: 083055248481416, 1-121967267, 1-429329281, DGC19607785, HC3062UV01, HCIT1401012UW, HCIT4331279UV, PACS-IW-127, PACS-IW-219, PACS-IW-2770, PACS-IW-323, PACS-IW-421, RICHMONDU14767331, RICHMONDU15531177N1, X1347906 / X1347909, X1347906 / X1347909, YP315021, ZA2412PAC21 and ZA2533PAC1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2502-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2502-2021Class IIOngoingVoluntary: Firm initiated
Z-2503-2021Class IIOngoingVoluntary: Firm initiated