Z-2503-2021Class II
Centricity Universal Viewer and Viewer 6.0 -
GE Healthcare / initiated 2021-08-06 / Open, Classified / root cause: software
GE Healthcare, LLC began a recall of Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging... on . FDA classified it as Class II and gives the reason as "Image acquisition failures and synchronization failure with the Centricity Enterprise Archive". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2503-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Includes the following Model Numbers: 1. Model Number: 2090255-001; 2. Model Number: 2088026-723; 3. Model Number: 2102676-001; 4. Model Number: 2104867-044.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 462 (279 US; 183 OUS)
- Distribution
- AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.
- Product codes and lots
- 1. Model Number: 2090255-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1) System ID (UDI-DI if applicable): PMIO14194768, PMIO20084199, PACS-IW-231, CITRUSVAL20884363, CITRUSVAL20884364, CITRUSVAL20884365, PACS-IW-155, PRIMEHEAL02315681, PRIMEHEAL20501355, PARKVIEWC21200827 (840682103800), PARKVIEWC21200830 (840682103800), PACS-IWUVN2-441, QUEEN19630914, QUEEN19630915, QUEEN19630916T, QUEEN20553950, 203268UVSW, DELAWAREC01589220, PACS-IW-462UV, PACS-IW-462UVN1, PACS-IW-126, PACS-IW-1062, OFMC01819402, OFMC20791988, HGS01766912, HGS20584004 (840682103800), GA001PACTRL01, TRGI02180958, DRY01810490, DRY19670358, PACS-IW-101, RML20502096, PACS-IWUV-239, PACS-IW-00813, PHS01374810, 01722652SEAS, 02031921SEASN1, MARYLANDF17985489, MYRTLEHIL17713895, MYRTLEHIL01444483, MYRTLEHIL17713895, VT01766830, VT01977629, PACS-IW-2637, SRMC01982629, PACS-IWUV-568, 1-142022244, PACS-IWUV-1131, OGAA02373799NOD1, ALLG00558571 (840682103800), EXC02189547, EXC01584099, PACS-IW-574, PRIME HEAL19508102, PRIME HEAL19050922, PRIME HEAL19050921, PACS-IW-219, ANCL00558570, ANCL02157634, NSW01452088, NSW20080545, MARYLANDF17985489, PACS-IW-518UV (840682103800), MTGD02393053, MTGD00558454, PACS-IW-44, QUENTIN01970717, LAKESIDEM02256808, BMC12P00558644P1, BMC12T02267839, BMC12N01271417N1, PACS-IW-58, SMCO02039946PC (840682103800), BWH01301357, PACS-IW-349, PACS-IW-255, 1-357059471, CAROCOMMU02205222, CAROCOMMU01459013, CSPA01291589 (840682103800), CSPA02097167 (840682103800), PACS-IW-79, PACS-IW-93, 1-276423210, MFCI01448329, LSFH02032334, CORONAREG01300927, 951737IWUWWN1, CORONAREG02288447, 1-382753509, CITRUSVAL19618406, MGMC02288391, MGMC02398531, 1-473998727, BARRETTHO00558704, BARRETTHO17737529, FORTDUNCA01298556 (840682103800), FORTDUNCA02395799 (840682103800), 1-283597082, TTU01973509, AVHL02396030, PACS-IW-511, PACSIWTEST, PACS-IW-386UV (840682103800), PACS-IW-486-UV (840682103800), PACS-IW-486-UVN1 (840682103800), PACS-IW-390 (840682103800), PACS-IW-202TEST, PACS-IW-202, HZC14961014, AVHL02396026, AVHL02396027, AVHL02396028, AVHL02396029, PACS-IW-419, PMRM14768587UVProxy, HANNIBALR02179353 (840682103800), TESTHANNIBALR02255613 (840682103800), HANN16174584UVProxy (840682103800), PACS-IW-189 (840682103800), HARTSDALE14768195 (840682103800), HOA01451973 (840682103800), RML01448365 (840682103800), ICA01367198 (840682103800), ICA14776604 (840682103800), JKS01452133 (840682103800), JKS14776447UVSFNode (840682103800), JKS14776448UVProxy (840682103800), JKS14776449UVTest (840682103800), PACS-IW-181, GTEC01477448, GTEC01477447, GTEC15993728UVProxy3, 1-592001096 (840682103800), UHSD14767800 (840682103800), PACS-IW-551UVN1 (840682103800), PACS-IW-551UVN2 (840682103800), UV1-154550926, DHOL19603374, DHOL19603375, PACS-IW-551UV (840682103800), PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, SFCS19667347, SFCS15961473, SFCS15961473, MHOG01806096, PACS-IW-166, PACS-IW-166T, MCT02277843, MCT02278145, PACS-IW-565, PACS-IWT-565, WGH00558334CTRL, WGH02372377PROX, WGH02372375BACK, 1-464783615, PACS-IW-389, BVH00558643, PACS-IW-348UV (840682103800), NRVM19609993UVT (840682103800), NRVM19611189UVPROXY (840682103800), PACS-IW-441, PACS-IW-441TUVSW, PACS-IW-402, PACS-IW-318, MINNETOHE, PACS-IW-253, PACS-IW-581UV, PACS-IW-370, PACS-IW-UVT370, VISOU02345023, VISOU02288781, 1-218291079, 1-103108035, MFHO14767332, PALMDALEM00558700, PALMDALEM01810788, PACS-IW-454, PACS-IW-406, 20101378UBSW, ILCC01373579 (840682103800), ILCC14580319 (840682103800), 516473UVSW (840682103800), JOHNTMA19732280T (840682103800), 1-182669905, NCDC20279342, CENTERFOR01627896, CENTERFOR16023960UVProxy, PACS-IW-176, PACS-IW-176UVTEST, PACS-IW-405, PACS-IW-336, NASONHOSP14898835, RRA01814001, MYRTLEHIL01444483, PSPC01815255, 1-283596904, HOOO20091140, GZMG02182469CNTRLR (840682103800), GZMG02287985 (840682103800), GZMG02288300 (840682103800), PACS-IW-556UV, URSA20025013, PACS-IW-431, 1-598891598, PACS-IW-230, INTERCOAST17594527UVProxy, MDHC16204735UVProxy, 201392UVSW, 505368UVBSW (840682103800), NNAR14767799 (840682103800), PACS-IW-479, PACS-IW-557 (840682103800), URS20583582 (840682103800), PACS-IWUV-481, PACS-IW-476, PACSIW476UVTEST, GGMC02349137, 1-486091448, 1-486091475, 1-750710672, SUC00558667, 1-802428413, 954688UVBSW100, PACS-IW-1511, KDH217192544, ACP00558109, PACS-IW-31Test1, PACS-IW-31Test, ACP01613226, 1-219095008, OKOS14767330, 918277CVUWBSW, CCA20759776SF, PACS-IW-487, QTHBL14767810UVPROXY, QTHBL14767809UVSFBK, QTHBL14767808UVSF1KO, QTHBL14767807UVSFSIM, QTHBL14767806UVSFVAL, QTHBL02256595, VCM02255583T, VCM01819618, VCM02255586TSP, VMC02255587NSP, BOJC00558520, PACS-IW-56, PACS-IW-225, 1-375705893, DDNI14776603, PACS-IW-129, DISSFHW01, 1-644600677, PACS-IW-154UV, FLMC14777071, NEWYORK01291738, OANT02035061, PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, VBHCP20497227, PACS-IW-2635, PREMI02182462, PREMI02182462, NORTHWEST01804127, HCIT4271649UW, AH4325SV01, AH5291PC02, AP6264PC01, AP6469PC01, AH0116SV03, AH0106SV03, HCD4440494UV, HCIT4258559UV, HCIT7109403UV, HCIT4244609UV, HCIT4708563UV, HCIT4470563UV-CTRL, HCIT4261024UVCTRL, HCIT7113573UV, 519264PACSAW1 (840682103800), 416813UV (840682103800), 416813TESTRPACSUV (840682103800), HCIT4266470UW, HCIT4238540UV, HCIT4486723UV, E4167658, X314119802, X141385617, A5485262 / A5485266 / A5485267 / A5485269 / A5485270 / A5485271, A9190177, M41980107, M4487535, M44997011, HC0399UV01, CC4105PC02, PC1695SV01, 786787SV02, HBON01SV02, HC4105PC01, 509940UV01, HC0529PC01, HC0529PC01, 233638UV01, HC0245PC02, 600791PC02, 427990UV01, PC1961SV02, 531344PC01, 434487SV03, PX4900SV02, HC0565PC03, TP0575UV02, GH1077UV01, 083055820116113, 083055824999515, 083055861021012, 83055862088813, 083055800046112, 81111DBS01, IL1469IT05, IL1469IT16, A5992711, A5412010, A6197102, KW0000PP02, HCIT4175024UV, AMS3963, BON0164, ECT0265, HCIT4192962UV, HCIT4264206UW, PL3673RW01, PL1493RW02, PL2970RW01, PL1124RW01, PL1499RW03, PL3349PC01, PL1501RW03, PL2624RW01, PL3249RW02, PL3277RW01, PL1005RW01, 600062UV, PL3362RW01, PL0326RW01, PL2975RW02, PL1504RW02, PL2971RW01, RU9255ID01, RU6968ID01, RU8945ID01, RU7519ID01, RU1781RW01, RU2580ID01, RU4255ID01, RU0003ID01, RU0472ID03, RU9595ID02, RU9327ID02, RU1011ID05, RU8366ID01, RU7808ID02, RU1516ID03, RU1646ID01, RU1112ID02, 01977813, 01980259, RU3052ID01, RU1107RW01, RU1107RW01, RU7175ID01, RU1717ID04, RU9216ID02, RU3453RW03, 600101RW02, RU8002ID01, RU1544ID01, RU1516ID03, RU9940RW01, RU5388ID01, RU0994ID03, RU3794ID01, RU9226ID01, RU9697ID01, 1509505, RU0968ID04, RU0120ID01, RU6164ID01, RU9546ID01, RU7468ID01, RU3998ID01, 600105ID03, SA2402PP02, SA2110PP05, SA1339PP03, ZA2541UV01, ZA2414PAC1, ZA2415PAC1, ZA2474UV01, ZA2267PAC1, ZA1058UV01, ZA2551PAC1, ZA1058UV01, ZA2412PAC2, ZA2472PAC2, ZA000010, ZA2412PAC2, MPX35219, DP760316, FP502804, 802030PC01, 228493WS01, 124774CS01, 482896SV03, 06051PAC04, HCIT4133734PACS, HCIT4148106PEE, Others Tracked under Service Cloud product ID. 2. Model Number: 2088026-723 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: NYHOSPI16204605, NYHOSPI16204606, NYHOSPI16204607, NYHOSPI16204608, NYHOSPI16204609, NYHOSPI16204612, NYHOSPI16204613. 3. Model Number: 2102676-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: PACS-IW-119. 4. Model Number: 2104867-044 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: DDOC02339896.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2502-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2503-2021 | Class II | Ongoing | Voluntary: Firm initiated |