Z-2476-2021Class II

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform...

Philips North America LLC / initiated 2021-08-13 / Terminated / root cause: design

Philips North America Llc began a recall of The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and... on . FDA classified it as Class II and gives the reason as "Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2476-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Root cause tag
design
FDA root cause
Device Design
Action
Philips notified accounts by Urgent Medical Device Recall letter via certified mailing United States Postal Service (USPS) on 8/13/21. Letter states reason for recall, health risk and action to take: Please place this Medical Device Recall letter with the documentation of the system until this correction has been implemented in your system. Additionally, please return the attached reply form to Phhilps to confirm that the users of the system have reviewed and understood this Medical Device Recall. Philips will correct the affected systems by replacing the ADUs. A Philips service representative will contact you to schedule the correct ion of your system. Please be assured that maintaining a high level of safety and quality is our highest priority. lf you need any further information or support concerning thls issue, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200483
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the states of ID, MA, MI, TX, VA.
Product codes and lots
Serial number: 921

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2476-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2474-2021Class IITerminatedVoluntary: Firm initiated
Z-2475-2021Class IITerminatedVoluntary: Firm initiated
Z-2476-2021Class IITerminatedVoluntary: Firm initiated