EVIS EXERA BF-XT160 Bronchofibervideoscope, Model No. BF-XT160
Olympus Corporation of the Americas began a recall of EVIS EXERA BF-XT160 Bronchofibervideoscope, Model No. BF-XT160 on . FDA classified it as Class II and gives the reason as "A gluing step was not performed during the manufacturing of the affected device's insertion tube. There is a potential for the insertion tube to develop a leak, which poses an infection control risk". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2497-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- EVIS EXERA BF-XT160 Bronchofibervideoscope, Model No. BF-XT160
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- A gluing step was not performed during the manufacturing of the affected device's insertion tube. There is a potential for the insertion tube to develop a leak, which poses an infection control risk.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- On August 17, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that Olympus will repair your unit free of charge. An Olympus representative will contact you to make arrangements for return of your affected bronchoscope. If you have further questions about this recall, please contact [email protected] for additional information.
- Quantity in commerce
- 27 globally (21 US)
- Distribution
- US Nationwide distribution in the states of CA, IL, MA, MD, MI, NY, OH, OR, and PA.
- Product codes and lots
- 1200148 2500873 2100707 2710983 1500300 1400209 2600505 1300163 1400273 1400248 2700521 2500431 1500310 1200129 2300805 2811029 1500292 2700520 2100743 1500313 2911060
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2497-2021 | Class II | Terminated | Voluntary: Firm initiated |