SARS-CoV-2 Process Control (Pellet) -
Microbiologics Inc began a recall of SARS-CoV-2 Process Control (Pellet) - Product Usage: intended for use as an external, non-viable control material to evaluate the performance of nucleic acid... on . FDA classified it as Class II and gives the reason as "The SARS-CoV-2 Process Control (Pellet) contains RNA transcripts that include diagnostically relevant SARS-CoV-2 targets (CDC and WHO consensus sequences)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2486-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- SARS-CoV-2 Process Control (Pellet) - Product Usage: intended for use as an external, non-viable control material to evaluate the performance of nucleic acid tests that detect SARS-CoV-2 virus.
- Product code
- OHQ Multi-Analyte Controls Unassayed
- Reason for recall
- The SARS-CoV-2 Process Control (Pellet) contains RNA transcripts that include diagnostically relevant SARS-CoV-2 targets (CDC and WHO consensus sequences).
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- On AUGUST 23, 2021 THE FIRM SENT THE FOLLOWING INSTRUCTIONS TO THEIR CONSIGNEES.: INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage. 3. COMPLETE the response form provided. 4. RETURN the response form to [email protected] 5. KEEP this letter for your records. 6. CONTACT Microbiologics to discuss possible alternatives if replacement is needed Microbiologics is dedicated to ensuring the success of your quality control processes. Please contact our Recall Support team at 320.229.7073 or [email protected] with questions or concerns. Collect calls may be made.
- Quantity in commerce
- 218
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IL, MD, MT, MN, NC, NY, NV, OR, TN, TX,UT, WA and WI. The countries of Austria, Canada, Colombia, France, Germany, Hong Kong, India, Ireland, Italy, Japan and United Kingdom.
- Product codes and lots
- Catalog # HE0062S Lot #'s: HE0062-01-1, HE0062-01-2, HE0062-03-1, HE0062-04-1, HE0062-04-2, HE0062-05-1, HE0062-05-2, HE0062-06-1 and HE0062-06-2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2486-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.