Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2487-2021Class II

SARS-CoV-2 Process Control (Swab) -

Microbiologics, Inc. / initiated 2021-08-20 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of SARS-CoV-2 Process Control (Swab) - Product Usage: intended for use as an external, non-viable control material to evaluate the performance of nucleic acid... on . FDA classified it as Class II and gives the reason as "The product indicated is contaminated with a plasmid DNA that contains the target SARS-CoV-2 sequences". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2487-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
SARS-CoV-2 Process Control (Swab) - Product Usage: intended for use as an external, non-viable control material to evaluate the performance of nucleic acid tests that detect SARS-CoV-2 virus.
Product code
OHQ Multi-Analyte Controls Unassayed
Reason for recall
The product indicated is contaminated with a plasmid DNA that contains the target SARS-CoV-2 sequences.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
On AUGUST 23, 2021 THE FIRM SENT THE FOLLOWING INSTRUCTIONS TO THEIR CONSIGNEES.: INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage. 3. COMPLETE the response form provided. 4. RETURN the response form to [email protected] 5. KEEP this letter for your records. 6. CONTACT Microbiologics to discuss possible alternatives if replacement is needed Microbiologics is dedicated to ensuring the success of your quality control processes. Please contact our Recall Support team at 320.229.7073 or [email protected] with questions or concerns. Collect calls may be made.
Quantity in commerce
127
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IL, MD, MT, MN, NC, NY, NV, OR, TN, TX,UT, WA and WI. The countries of Austria, Canada, Colombia, France, Germany, Hong Kong, India, Ireland, Italy, Japan and United Kingdom.
Product codes and lots
Catalog # HE0063S Lot #'s: HE0063-100-1, HE0063-103-2, HE0063-103-3, HE0063-103-1, HE0063-104-1, HE0063-101-1 and HE0063-100-1.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2487-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2487-2021Class IITerminatedVoluntary: Firm initiated