MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key...
Siemens Medical Solutions USA, Inc began a recall of MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815 on . FDA classified it as Class II and gives the reason as "Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2447-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens has initiated three (3) Customer Safety Advisory Notices (CSANs) to affected customers via SY052/21/S, CT042/21/S and MR019/21/S by email or hand delivery by the Siemens Customer Service Engineer on 8/16/21. Temporary workarounds are available for some of the software versions and are described in the CSANs. Additionally, a solution to eliminate the root cause of this problem is being distributed via SY051/21/S (for SY and CT) and MR021/21/S for MR to all affected customers free of charge. The service organization will contact you to arrange a date to perform this corrective action free of charge. Contact the service organization for an earlier appointment at 1-800-888-7436. ***UPDATE: 01/31/22 Additionally, a solution to eliminate the root cause of this problem is being distributed via CT049/21S, CT061/21/S, CT062/21/S, and CT088/21/S for CT. ***UPDATE: 03/12/22 Status Report Additional 13 S/N's added
- Quantity in commerce
- 2 systems
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Serial Numbers: 176302 175681
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2445-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2446-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2447-2021 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.