MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key...

Siemens Medical Solutions USA, Inc. / initiated 2021-08-16 / Open, Classified / root cause: software

Siemens Medical Solutions USA, Inc began a recall of MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815 on . FDA classified it as Class II and gives the reason as "Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2447-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Root cause tag
software
FDA root cause
Software design
Action
Siemens has initiated three (3) Customer Safety Advisory Notices (CSANs) to affected customers via SY052/21/S, CT042/21/S and MR019/21/S by email or hand delivery by the Siemens Customer Service Engineer on 8/16/21. Temporary workarounds are available for some of the software versions and are described in the CSANs. Additionally, a solution to eliminate the root cause of this problem is being distributed via SY051/21/S (for SY and CT) and MR021/21/S for MR to all affected customers free of charge. The service organization will contact you to arrange a date to perform this corrective action free of charge. Contact the service organization for an earlier appointment at 1-800-888-7436. ***UPDATE: 01/31/22 Additionally, a solution to eliminate the root cause of this problem is being distributed via CT049/21S, CT061/21/S, CT062/21/S, and CT088/21/S for CT. ***UPDATE: 03/12/22 Status Report Additional 13 S/N's added
Quantity in commerce
2 systems
Distribution
US Nationwide distribution.
Product codes and lots
Serial Numbers: 176302 175681

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2447-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2445-2021Class IIOngoingVoluntary: Firm initiated
Z-2446-2021Class IIOngoingVoluntary: Firm initiated
Z-2447-2021Class IIOngoingVoluntary: Firm initiated