Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330
Biodex Medical Systems, Inc. began a recall of Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330 on . FDA classified it as Class II and gives the reason as "When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2580-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sport Medical Technology Inc
- Product
- Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330.
- Product code
- KPT Calibrator, Dose, Radionuclide
- Reason for recall
- When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Biodex issued Urgent Medical Deice Recall Letter on 8/16/21 via email. Letter states reason for recall, health risk and action to take: Review the dial settings associated with custom isotope and confirm they have not been changed. Compare the measured activity with a dose calibrator from another manufacturer The error can be immediately corrected by restarting the system. We recommend that you promptly download new software. The error is corrected in Software Revision 2.0.10. The corrected software upgrade is available as a download from the Biodex website at the following link Atomlab 500 Dose Calibrator Software Update v2.0.10 - Software Updates - Nuclear Medicine | Biodex. System registration is required for download. Please email us at [email protected] to acknowledge receipt of this letter and confirm commitment to download the software update. To minimize risk of illness or injury, please download the software correction, as soon as possible. If you have any questions or require any additional information, please contact [email protected].
- Quantity in commerce
- 829 units
- Distribution
- Nationwide Foreign: Austria Autralia Brazil CANADA Chile Colombia Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt France Hong Kong HUNGARY Iceland India Iraq Ireland ISRAEL Isreal Italy Japan JORDAN Kuwait Lebanon Libya Malaysia MALTA Mexico Morocco Netherlands New Zealand Oman Pakistan Palestine Panama Peru Philippines Poland Portugal PR Puerto Rico Qatar Romania Russia Saudi Arabia Singapore South Africa South Korea Spain Sweden Taiwan UAE United Arab Emirates United Kingdom Venezuela
- Product codes and lots
- Serial Number range:17081557-21062362 UDI: 00718175003305
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2580-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2581-2021 | Class II | Terminated | Voluntary: Firm initiated |