Z-2508-2021Class II
Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2...
Siemens Healthcare Diagnostics Inc. / initiated 2021-08-05 / Terminated
Beckman Coulter Inc. began a recall of Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2 IgG (1st IS) Calibrators, REF C74340 Access... on . FDA classified it as Class II and gives the reason as "Firm's submission for Emergency Use Authorization (EUA) contained insufficient evidence to be granted EUA Authorization to distribute". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2508-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access SARS CoV-2 IgG (1st IS) Assay: Access Immunoassay Systems REAGENT PACK, REF C74339 Access SARS-CoV-2 IgG (1st IS) Calibrators, REF C74340 Access SARS-CoV-2 (1st IS) Quality Control (QC), REF C74341
- Product code
- QKO Reagent, Coronavirus Serological
- Reason for recall
- Firm's submission for Emergency Use Authorization (EUA) contained insufficient evidence to be granted EUA Authorization to distribute
- FDA root cause
- No Marketing Application
- Action
- On 08/06/2021, the firm sent an initial "URGENT MEDICAL DEVICE RECALL" Notification informing its customers that due insufficient evidence the firm was not issued an Emergency Use Authorization for its SARS-CoV-2 IgG (1st IS) assay and required to discontinue distribution and to recall all affected products. On 09/13/2021 the firm sent a revised "URGENT MEDICAL DEVICE RECALL UPDATE" to inform its customers about the potential risks/harms related to using the affected products, such as, the assay may generate discordant results such as false positive, false negative or misinterpreted results when compared to other World Health Organization (WHO) traceable assays. Customers are instructed to: 1. Immediately discontinue using the Access SARS-CoV-2 IgG (1st IS) assay. 2. Discard all Access SARS-CoV-2 (1st IS) reagent packs, calibrators, and quality control. 3. A retrospective review of patient test results obtained with the Access SARS-CoV-2 IgG (1st IS) assay is not required. Any review is left to the discretion of the Laboratory Medical Director. 4. Share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. 5. If any of the affected products have been forwarded to another laboratory, provide them with the notification. 6. Complete and return the Customer Response Form within 10 days electronically via email or manually via mail 7. For credit for the affected products, complete the Customer Response Form For questions, contact Customer Support Center via: Website - http://www.beckmancoulter.com Phone - 1-800-854-3633 in the United States
- Quantity in commerce
- 303 kits (U.S.) and 964 kits (O.U.S.)
- Distribution
- U.S. Nationwide distribution in the states of AZ, CA, GA, IL, IA, KY, MI, NJ, NY, NC, OH, PA, Puerto Rico, TN, TX, VA, and WA.
- Product codes and lots
- Part Number: C74339 UDI Code: (01)15099590746919(17)210930(11)210402(10)124466 Lot Number: 124466; Part Number: C74340 UDI Codes: (01)15099590746926(17)210930(11)210402(10)124467; (01)15099590746926(17)211231(11)210703(10)124758; Lot Numbers: 124467; 124758; Part Number: C74341 UDI Codes: (01)15099590746933(17)210930(11)210402(10)124468; (01)15099590746933(17)211130(11)210602(10)124708; Lot Numbers: 124468; 124708;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2508-2021 | Class II | Terminated | Voluntary: Firm initiated |