Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA...
Philips Ultrasound Inc began a recall of Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only on . FDA classified it as Class II and gives the reason as "Battery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0070-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Battery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On August 24, 2021, Philips issued "Urgent Medical device Correction" notification to affected consignees. The customer communication was distributed via: USPS Certified Mail In addition to informing consignees about the recall, Philips requested consignees take the following actions: Instructions for how to determine the software version of your Ultrasound system: 1. Power up the system and allow it to complete the boot sequence, 2. Press the Wrench icon on the upper right side of the control panel, 3. Press the Service button on the bottom left corner, 4. Under System Management, Press System Information, 5. The software version is listed in the Software Information Section. 4. Describe the actions that should be taken by the customer / user in order to prevent risks for patients or users A. If this shutdown occurs, Philips recommends that the system be put back on AC power and restarted. The system will be fully functional upon restart. However, restarting the system will not prevent the possibility of the issue reoccurring. B. Please complete and return the attached form to Philips promptly and no later than 30 days from receipt (via email: [email protected], or fax: 1-833-512-7756). 5. Describe the actions planned by Philips Ultrasound to correct the problem A Philips Field Service Engineer will contact you to schedule a software update to permanently resolve the issue (reference FCO79500550). 6. This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. 7. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 and reference FCO79500550.
- Quantity in commerce
- 34 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, MA, MD, MI, MS, NC, NY, OH, VA, WA and the countries of Australia, Portugal, Netherlands, Canada, Netherlands, Germany.
- Product codes and lots
- Affected Serial Numbers/UDI numbers: USD1410130 /N/A, US12110905 /(01)00884838097971(21)US12110905, US12110906 /(01)00884838097971(21)US12110906, US12111049 /(01)00884838097971(21)US12111049, US31310191 /N/A, US31511242 /N/A, US32111072 /(01)00884838097971(21)US32111072, US41210049 /N/A, US42111774 /(01)00884838097971(21)US42111774, US51210186 /N/A, US62011642 /(01)00884838028098(21)US62011642, US71810505 /(01)00884838028098(21)US71810505, US72011713 /(01)00884838028098(21)US72011713, US72011714 /(01)00884838028098(21)US72011714, US82010111 /(01)00884838028098(21)US82010111, US91410757 /N/A, US92010327 /(01)00884838028098(21)US92010327, US92010512 /(01)00884838028098(21)US92010512, USD2011313 /(01)00884838097971(21)USD2011313, USD2011314 /(01)00884838097971(21)USD2011314, USD2011362 /(01)00884838097971(21)USD2011362, USD2011363 /(01)00884838097971(21)USD2011363, USD2011480 /(01)00884838097971(21)USD2011480, USD2011481 /(01)00884838097971(21)USD2011481, USD2011482 /(01)00884838097971(21)USD2011482, USD2011483 /(01)00884838097971(21)USD2011483, USD2011578 /(01)00884838097971(21)USD2011578, USD2011622 /(01)00884838097971(21)USD2011622, USD2011623 /(01)00884838097971(21)USD2011623, USN1510828 /N/A, USN2011691 /(01)00884838097971(21)USN2011691, USO1210225 /N/A, USO2011257 /(01)00884838097971(21)USO2011257, USO2012052/ (01)00884838097971(21)USO2012052
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0070-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0071-2022 | Class II | Terminated | Voluntary: Firm initiated |