HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated...
Instrumentation Laboratory began a recall of HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH)... on . FDA classified it as Class II and gives the reason as "Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0149-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma on IL Coagulation Systems, (ACL TOP Family, ACL TOP Family 50 Seriest and ACL Elite/Elite Pro/8/9/10000) Part Number: 0020302600
- Product code
- KFF Assay, Heparin
- Reason for recall
- Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Werfen issued Urgent Medical Device Correction letter via email via email or registered mail on 9/7/21. Letter states reason for recall, health risk and action to take: NOTE: Please refer to the separate "URGENT Medical Device Removal" notice (dated August 25, 2021) for Hemosll Liquid Anti-Xa, Part No. 0020302600, which removes claims for this assay from the ACL Elite and ACL Elite Pro instruments. " Mandatory Customer Actions At this time, the on-board instrument stability claim has been reduced from 7 days to 5 days for all in-date and future product lots of Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602) on the ACL TOP Family and ACL TOP Family 50 Series. Please take the following immediate actions: " Use the following reduced on-board instrument stability claim for Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602) on the ACL TOP Family and ACL TOP Family 50 Series. ACL TOP Family and ACL TOP Family 50 Series Current On-board Instrument Stability Reduced On-board Instrument Stability 7 Days 5 Days " Run quality controls before patient testing or every 8 hours and with each new vial in accordance with good laboratory practice. " Post this notification on each of your ACL TOP Family and ACL TOP Family 50 Series instruments. " Share this information with your staff, notifying them of the reduced on-board stability requirement of 5 Days for the ACL TOP Family and ACL TOP Family 50 Series. " Retain a copy of this letter in your files as a record of the notification. " Complete and return the enclosed Mandatory Response Tracking Form to acknowledge the above actions. " Return the completed and signed form immediately to: " Fax No.: 781-861-4207 " E-mail: [email protected] NOTE: Please return the tracking form even if you do not have any inventory of Hemosll Liquid Anti-Xa (Part Nos. 0020302600 and 0020302602). " Contact Information for Questions " For technical questions, please contact the Technical Support Center at 1-8
- Quantity in commerce
- 64,114
- Distribution
- US Nationwide Foreign: Canada Armenia Austria Australia Belgium Brunei Brazil Belarus Switzerland China Colombia Czech Republic Germany Denmark Spain Finland France United Kingdom Gibraltar Greece Hong Kong Hungary Indonesia Ireland Israel India Iran Italy¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ Japan South Korea Kuwait Kazakhstan Lebanon Sri Lanka Lithuania Luxembourg Macau Mexico Malaysia Netherlands Poland Portugal Qatar Reunion Romania Serbia Russia Saudi Arabia Sweden Thailand Tunisia Turkey Taiwan Vietnam Zambia
- Product codes and lots
- All currently released lots and future lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0149-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0150-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0151-2022 | Class II | Ongoing | Voluntary: Firm initiated |