Z-0109-2022Class II

Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal...

Draeger, Inc. / initiated 2021-09-10 / Open, Classified / root cause: software

Draeger Medical, Inc. began a recall of Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs)... on . FDA classified it as Class II and gives the reason as "Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0109-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Draeger, Inc.
Product
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Root cause tag
software
FDA root cause
Software design
Action
Drager issued Urgent Medical Device Recall letter on 9/10/21 via FedEx with tracking (confirmation of delivery). Letter states reason for recall, health risk and action to take: The release of a corrected software is expected to be available in Q4 of 2021. Once the software is available, your local Drger Service representative will contact you to arrange a date for the software update to be performed free of charge. According to our records, at least one of the mentioned affected devices has been shipped to your facility. Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Recall Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us that you have received this information. If you have any questions regarding the operation of your Evita V500 or Babylog VN500, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1).
Quantity in commerce
2,169 units
Distribution
US Nationwide
Product codes and lots
Software 2.51 and Lower

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0109-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0108-2022Class IIOngoingVoluntary: Firm initiated
Z-0109-2022Class IIOngoingVoluntary: Firm initiated