Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83...
Datascope Corp. began a recall of Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85 on . FDA classified it as Class II and gives the reason as "The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2616-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Datascope Corp.
- Product
- Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On Wednesday, September 8, 2021, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that all lots of Maquet Cardiosave Hybrid experience the following issue: Unexpected Shutdown Upon Battery Removal. Immediate Actions to be taken by Customers: To ensure that the Cardiosave Hybrid is not susceptible to shutdown during the removal of a battery, users should not eject a battery when: - There is a singulat battery in either battery charging bay and unused battery bay is empty. - The battery is charging on AC power (indicated by a flashing LED on the battery pack). Additionally, the user can prevent the issue by ensuring that a battery is inserted into each of the two battery bays. Should a charging battery be ejected inadvertently, the unit will automatically switch over to the second battery, thus preventing system shutdown. Furthermore, Datascope/Getinge is developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
- Quantity in commerce
- 3969 total (US); 3550 total (OUS)
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2615-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2616-2021 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.