Z-0140-2022Class II

Optima MR450w 1.5T nuclear magnetic resonance imaging system

GE Healthcare / initiated 2021-08-26 / Terminated / root cause: software

GE Healthcare, LLC began a recall of Optima MR450w 1.5T nuclear magnetic resonance imaging system on . FDA classified it as Class II and gives the reason as "Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0140-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Optima MR450w 1.5T nuclear magnetic resonance imaging system
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Root cause tag
software
FDA root cause
Software design
Action
GE issued an Urgent Medical Device Correction letter to customers with affected devices. This letter informs the customer of the issue and provides safety instructions for the continued use of the device. GE Healthcare will correct all affected devices at no cost to the customer. In the U.S., this letter will be sent by a traceable means. This letter will be sent to the following titles within the affected accounts: Director of Clinical/Radiology, Risk Manager/Hospital Administrator.
Quantity in commerce
13 devices
Distribution
Distribution in United States to AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH, NJ, NY, OH, OK, TX, VA, WA, and Puerto Rico OUS distribution to: Australia, Belgium, Canada, Chile, Colombia, France, Germany, Italy, Japan, Korea, Poland, Saudi Arabia, Spain, Sweden, Turkey, UAE, and UK
Product codes and lots
Software version DV29.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0140-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0132-2022Class IITerminatedVoluntary: Firm initiated
Z-0133-2022Class IITerminatedVoluntary: Firm initiated
Z-0134-2022Class IITerminatedVoluntary: Firm initiated
Z-0135-2022Class IITerminatedVoluntary: Firm initiated
Z-0136-2022Class IITerminatedVoluntary: Firm initiated
Z-0137-2022Class IITerminatedVoluntary: Firm initiated
Z-0138-2022Class IITerminatedVoluntary: Firm initiated
Z-0139-2022Class IITerminatedVoluntary: Firm initiated
Z-0140-2022Class IITerminatedVoluntary: Firm initiated
Z-0141-2022Class IITerminatedVoluntary: Firm initiated