Z-0031-2022Class II

Nox T3s Recorder -

Nox Medical / initiated 2021-09-07 / Terminated / root cause: software

NOX MEDICAL began a recall of Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep on . FDA classified it as Class II and gives the reason as "Error in the device firmware results in the inaccurate detection of device position". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0031-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nox Medical
Product
Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.
Product code
MNR Ventilatory Effort Recorder
Reason for recall
Error in the device firmware results in the inaccurate detection of device position.
Root cause tag
software
FDA root cause
Finished device change control
Action
Written notification letter titled, "URGENT: Medical Device Recall" was sent to all consignees on 09/07/2021. The letter instructs the consignee to follow the following instructions: "Further distribution or use of the affected devices should cease immediately. Devices should be quarantined pending device firmware update, per the instructions below. Due to the potential of inaccurate position-related reporting of TST, TRT, AHI and REI; studies performed with the affected T3s devices should be reanalyzed to determine if this has clinical impact on the patient diagnoses. A Nox Medical technical representative will contact you within 5 business days to schedule a remote session (via videoconference or other means) to upgrade the device firmware. Nox Medical requests that you formally acknowledge receipt of this notice and return the attached Customer / Distributor - Acknowledgement and Receipt Form (Attachment 1). This recall notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If any affected devices which were in your possession (indicated in Attachment 2: Product Distribution and Information Table) have been transferred, please indicate this and provide as much detail as possible in the Acknowledgement and Receipt Form (Attachment 1)."
Quantity in commerce
128 units
Distribution
US Nationwide distribution in the states of OH, IL, GA, IA, CA, PA, NM, MI, SD.
Product codes and lots
Catalog Number: 561111; UDI: (01)15694311111443; Serial Numbers: 300106423, 300106426, 300106420, 300106419, 300106422, 300106427, 300106425, 300106424, 300106421, 300106477, 300106485, 300106783, 300106806, 300106841, 300106843 , 300106846, 300106848, 300106852, 300106853, 300106857, 300106862, 300106881, 300106882, 300106883, 300106884, 300106885, 300106886, 300106888, 300106889, 300106890, 300106894, 300106897, 300106912, 300107004, 300106385, 300106396, 300106397, 300106407, 300106413, 300106414, 300106399, 300106415, 300106416, 300106475, 300106476, 300106483, 300106486, 300106494, 300106534, 300106537, 300106538, 300106544, 300106545, 300106547, 300106548, 300106552, 300106648, 300106658, 300106675, 300106677, 300106678, 300106684, 300106754, 300106772, 300106775, 300106807, 300106809, 300106810, 300106812, 300106839, 300106842, 300106849, 300106855, 300106868, 300106891, 300106892, 300106899, 300106909, 300106926, 300106951, 300107003, 300106398, 300106481, 300106542, 300106546, 300106782, 300106910, 300106400, 300106535, 300106536, 300106808, 300106950, 300106469, 300106470, 300106869, 300106854, 300106478, 300106805, 300106811, 300106840, 300106850, 300106856, 300106859, 300106887, 300106893, 300106895, 300106896, 300106898, 300107005, 300106776, 300106781, 300106789, 300106788, 300106833, 300106549, 300106543, 300106557, 300106540, 300106533, 300106556, 300106555, 300106901, 300106786, 300106553, 300106784, 300106541, 300106777, 300106774

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0031-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0031-2022Class IITerminatedVoluntary: Firm initiated