AngioDynamics NanoKnife Disposable Single Electrode Probes, Irreversible Electroporation (IRE) Device, 15 cm...

Boston Scientific Corporation / initiated 2020-04-24 / Terminated / root cause: software

Angiodynamics, Inc. began a recall of AngioDynamics NanoKnife Disposable Single Electrode Probes, Irreversible Electroporation (IRE) Device, 15 cm IRE Single Electrode RFID Activation, PG. Catalog... on . FDA classified it as Class II and gives the reason as "Programming issue affected RFID function of a single lot and did not allow the NanoKnife probes to be recognized by the NanoKnife generator". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0073-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
AngioDynamics NanoKnife Disposable Single Electrode Probes, Irreversible Electroporation (IRE) Device, 15 cm IRE Single Electrode RFID Activation, PG. Catalog No. 20400103
Product code
OAB Low Energy Direct Current Thermal Ablation System
Reason for recall
Programming issue affected RFID function of a single lot and did not allow the NanoKnife probes to be recognized by the NanoKnife generator.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
On April 24, 2020, the firm distributed recall notifications to their consignees, who were all international (non-US). Customers were advised to stop using the recalled product, remove it from inventory, and segregate in a secure location for return. The firm provided further instructions for the return of the product.
Quantity in commerce
37
Distribution
Internationally distribution to the countries of Australia, Germany, Denmark, France, and Thailand.
Product codes and lots
lot 5577754

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0073-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0073-2022Class IITerminatedVoluntary: Firm initiated