Z-0086-2022Class II
Endoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody...
Medtronic Sofamor Danek USA Inc / initiated 2021-09-02 / Terminated / root cause: packaging
Medtronic Sofamor Danek USA Inc began a recall of Endoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody, Medium, 10mm on . FDA classified it as Class II and gives the reason as "Engraving on implants may not match what is listed on the pouch labels". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0086-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Endoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody, Medium, 10mm
- Product code
- OVE Intervertebral Fusion Device With Integrated Fixation, Cervical
- Reason for recall
- Engraving on implants may not match what is listed on the pouch labels.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- A customer letter was sent via FedEx to impacted US consignees and delivered on or about September 3, 2021. Customers are asked to locate and remove the impacted product from normal storage locations. Customers are advised to not use product. A Customer Confirmation Form should be completed and returned to [email protected]. Customers may also coordinate return and replacement of impacted product with the Medtronic Sales Representative. This recall notice should be forwarded to those within the customer organization or to any organization where potentially impacted devices have been transferred.
- Quantity in commerce
- 9 units
- Distribution
- Distribution in US: CA, MN, MO, NC, NV, NY, OH, OK, OR, PA, SC, SD, WA, WV
- Product codes and lots
- Product Number 5366-1410-N, Lot A200510 GTIN 00191375018490
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0085-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0086-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.