Medtronic Bio-Medicus Insertion Kits, Model number 96553, Pediatric Insertion Kit, Sterile; and Model number...
Medtronic Perfusion Systems began a recall of Medtronic Bio-Medicus Insertion Kits, Model number 96553, Pediatric Insertion Kit, Sterile; and Model number 96551, Venous Insertion Kit, Sterile on . FDA classified it as Class II and gives the reason as "There was a label mix-up between two model numbers resulting in the incorrect guidewire being contained in the kits for both model numbers". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0003-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Medtronic Bio-Medicus Insertion Kits, Model number 96553, Pediatric Insertion Kit, Sterile; and Model number 96551, Venous Insertion Kit, Sterile.
- Product code
- DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Reason for recall
- There was a label mix-up between two model numbers resulting in the incorrect guidewire being contained in the kits for both model numbers.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm issued letters dated September 2021 beginning 9/13/2021 via 2-day UPS delivery to consignees. For outside of the United States, the recall letter will be delivered via regionally-approved methods to consignees who received the product. The letter explains the issue and requests the consignee review their inventory for the affected product. The product is to be immediately quarantined and returned after first contacting Customer Service. A Customer Confirmation Form was enclosed to be returned via email. The letter was to be shared with others within their organization as appropriate and to any other locations who have received the product through a transfer. Consignees are asked to review the Medtronic Urgent Medical Device Recall notification and follow the Recommended Actions, including quarantining and returning all unused affected Bio-Medicus Insertion Kits in their inventory.
- Quantity in commerce
- 168 kits
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of Massachusetts and the countries of Belgium, France, Germany Japan, Korea, Saudi Arabia, Spain, and Turkey.
- Product codes and lots
- Model 96553 - Serial number 220719040, GTIN 20763000116006; Model 96551 - Serial number 220641720, GTIN 20763000115986.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0003-2022 | Class II | Ongoing | Voluntary: Firm initiated |